SEEG 100
The EEG performance tester designed for regulation testing.
Standard Coverage: IEC 60601-2-26:2012, GB 9706.226-2021, JJG 1043, JJF 1388, JJF 1390.
Applicable for R&D, compliance and production line testing
OVERVIEW Designed for the latest EEG standard IEC 60601-2-26:2012 and GB 9706.226-2021 Sophisticated electrical and shielding design enables the test system to output 1μV signal 9 channels with auto-select switches for user to test performance individually Built-in test circuits as defined in EEG standards with selection of 620kΩ/4.7nF for input impedance test Able to load and replay files in PhysioNet's EDF format, binary format and text format IEC 60601-2-26 and GB 9706.226 helper simplifies the medical standard with test sequences, options, parameters and pass criterions Software Development Kit (SDK) assists user to develop customized or automated test software with less efforts